Salt spray tester test method

Salt spray tester test method

Clicks:383 Release time:2010/9/9

The salt spray tester test method can be set according to the standards of CNS, JIS, ASTM, etc.:
(A) Salt test method:
(1) Laboratory: 35 °C ± 1 °C
(2) Saturated air bucket: 47 ° C ± 1 ° C
(B) Corrosion resistance test method:
(1) Laboratory: 50 °C ± 1 °C
(2) Saturated air tank: 63 °C ± 1 °C
(C) Copper Accelerated Test Method: CASS
(1) Laboratory: 58 ° C ± 1 ° C
(2) Saturated air bucket: 65 °C ± 1 °C
Add glacial acetic acid to make the pH of the test solution 3.1~3.3, and match 0.02%~0.025% copper chloride