Introduction of high and low temperature impact environment testing machine

Two-box thermal shock test chamber, high and low temperature impact environment testing machine

The thermal shock test chamber, also known as the environmental test machine, tests various materials for heat resistance, cold resistance, dry resistance and moisture resistance. Suitable for electronics, electrical appliances, food, vehicles, metals, chemicals, building materials and other factories.

Control methods and features :

The thermal shock test chamber utilizes low temperature and high temperature cold storage hot tanks to move the basket quickly into the hot and cold tank according to the action requirements , so as to achieve rapid temperature impact effect , balance temperature control system ( BTC ) + special design air supply circulation system to the SSR PID controlled, so that the amount of heat equal to the amount of heat loss of the system, so that long-term stable use.

First, the model : thermal shock test machine   WDGJ-64

Second, performance : refers to air-cooled at room temperature + 25 °C When it is empty .

1. Test tank temperature range : - 40 °C -- + 100 °C .

2. High temperature bath temperature setting range : + 40 °C -- + 100 °C .

3. High temperature bath heating time RT (room temperature) -- + 100 °C It takes about 30min (room temperature is + 10 -- + 30 °C Time) .

4. Low temperature tank temperature setting range : - 40 °C -- - 10 °C .

5. Low temperature tank cooling time RT (room temperature) --   - 40 °C About 55min (room temperature at + 10 -- + 30 °C Time) .

6. Temperature fluctuation : ± 1.0 °C .

7. Temperature uniformity : ± 2.0 °C .

8. Impact return time: - 40 -- + 80 °C About need   5min. The conversion time is 8 seconds.

9. High and low temperature impact constant temperature time is more than 30min each

Third, the structure :

1. Internal dimensions :    W40 × H40 × D40 cm .

2. The external dimensions are approximately : W170 × H180 × D200 cm .

3. Inner box material : mirror stainless steel plate ( SUS #304 ) .

4. Outer box material : sand grain stainless steel plate ( SUS#304 )

5. Insulation material :  

a. High temperature tank: aluminum silicate insulation cotton .

b. Low temperature tank: high density PU foaming .

6. Box door : single door on the upper and lower boxes , left open .

a. Embedded plane handle .

b. Back button : SUS#304.

c. Silicone foam strips .

7. Sample placement rack : a. Basket size: W400 x D400 mm .

b. The weight of the basket is not greater than 5KG .

c. Stainless steel SUS#304 mesh basket storage tray 2 pieces .

Fourth, the heating system : fin-type heat pipe shaped stainless steel electric heater .

1. The high-temperature tank   8 KW.

2. Low temperature tank   5 KW.

Fifth, the air circulation system :

1. Motor 1HP × 2 sets .

2. Stainless steel lengthened shaft .

3. Multi-blade fan blades ( SIROCCO FAN ) .

4. Specially designed fan forced air circulation system .

Sixth, the freezing system :

1. Compressor : Fully enclosed high efficiency French Taikang Iron Armor .

2. Refrigerant : environmentally friendly refrigerant .

3. Condenser : air-cooled condenser .

4. Evaporator : multi-section automatic load capacity adjustment of fins .

5. Expansion system : automatic capacity control refrigeration system .

Seven, impact temperature controller:

LCD touch screen direct button type controller , in English and Chinese 5.7 ” Wide viewing angle of the graphics , high contrast LCD LCD controller with adjustable backlight function .

1. Controller specifications : a. Accuracy : temperature ± 0.1 °C + 1digit.

b. Resolution : Temperature ± 0.1.

c. With upper and lower limit standby and alarm function .

d. Temperature input signal   T type .

ePID control parameter setting , PID automatic calculation .

2.. Screen display function : a. Adopt screen-to-talk type , no need to input key , screen direct touch option .

b. Temperature setting ( SV ) and actual ( PV ) value are displayed directly .

c. Display the fault status and explain the troubleshooting method .

d. It can display the current execution program number , the number of times , the remaining time and the number of cycles .

e. The temperature program setting value is displayed in a graphical curve with real-time display program curve execution function .

f. With separate program editing screen , you can input temperature , time and cycle times .

g. The screen can be used for backlight adjustment .

h. The screen display protection function can be used for timing, TIMER or manual shutdown settings .

And 3. The capacity control program: a program group may be used:. Zui 96 PATTEN large (i.e., test specifications can be set to 96).

b. Repeatable execution of commands : up to 999 times per command .

c.SEGMENTS time setting 0 -- 99Hour59Min.

e. With power-off program memory , automatically start and resume the program function after power-on .

f. The graph curve can be displayed in real time during program execution .

g. With appointment start and shutdown function .

h. With date , time adjustment function .

 